Page 3 · Read text
Specifications
Measuring Range of Illuminance:
Low: 0 -1000 Lx (Red scale)
Ordinary: 0-5000 Lx (Black scale)
High: 0-50000 Lx (by using the Slide of x10 multiplier)
Accuracy : ± 10% (tested by a standard parallel-light tungsten lamp of color
temperature 2854°K) without including the resolution for
the meter (the value deserves minimal scale 1/2)in this
accuracy.
Angle Allowance : 30'---Iess than -3% 60°---less than -10%
Correction of sight sensitivity
0-1000 Lx (No correction-light source factor used
0-5000 Lx (Correction used)
Power Source : No battery used
Condition of Use : Temperature: 5 - 35° C Humidity: 45 -85
Accessories : Lumidisc for low measurement
Lumidisc for ordinary measurement
Slide ( x10 multiplier)
Dimensions : 112X58X27mm Weight : 135g
Characteristics : Compact, light weight, handy to carry, and most easy -to-use
swivel-head system
Name of
Parts
Donated to www.orphancameras.comView original page 3Page 4 · Read text
How to operate the Illuminometer
1. Getting ready for measurement
All you do for preparation is just to insert the LUMIDISC into position. To insert it,
match the white dot (or the red dot for low measurement) on the lumidisc to the white
dot on the swivel-head of the meter. Turn the lumidisc clockwise lightly to about 45°
until it tightly fits.
2. Ordinary measurement: 0--5000 Lx
To measure the range of 0--5000 Lx, use the lumidisc for ordinary measurement (with
a white dot at the outer ring of the disc and a green filter attached at the back).
Insert it into position as mentioned in 1. "Getting ready". Then place the meter at the
surface to be measured so that the lumidisc is parallel to the surface. Read the BLACK
figure on the scale. The figure the pointer indicates is the correct lux to use.
3. Measurement of High range: 0 --50000 Lx
To measure the range of 0--50000 Lx, use the lumidisc for ordinary measurement (see
2. above). Insert it into position in the same procedure as in 2. Then INSERT THE
SLIDE (x10 multiplier) into the slit located at the top of the swivel-head. Measure in
the same way as mentioned in 2. "Ordinary measurement". Read the BLACK figure the
pointer indicates, then MULTIPLY the figure by 10 to get the correct lux.
4. Measurement of Low range: 0--1000 Lx
To measure the range of 0-1000 Lx, use the lumidisc for low measurement (with a red
dot at the outer ring of the disc and no filter at the back). Insert it into position and
measure in the same way as in 2. Read the RED figure on the scale. Then MULTIPLY
the figure by the correction factor of each light source shown in the table below.
Correction factor Example
Mercury arc lamp...............................1.1 In case the pointer indicates 600 Lx
Fluorescent lamp ...............................1.0 under mercury arc lamp, the correct
Incandescent light..............................1.0 lux is, 600 Lx X 1.1-=660 Lx
Day light..........................................0.8
Donated to www.orphancameras.comView original page 4Page 5 · Read text
5. Swivel-Head
The swivel-head is one of the most advantageous features of this meter. Swivel the
head of the meter to the right or left when the measurement is difficult due to the
position of the surface to be measured.
6. Zero setting adjustment
When the light receptor is completely covered by hand (a condition where no light
enters into the light receptor), the pointer should be at zero. If it isn't, turn the zero
adjusting screw on the back until the pointer comes to the zero position.
7. General Precautions
1) Be careful not to give any shadow on the light receptor at every
measurement.
2) Position the meter (more exactly the surface of Lumidisc) in parallel to the
surface to be measured.
3) Avoid to leave the meter for long hours in a place of high temperature and
humidity.
4) Store it in the case and keep in a dry place when not in use.
SEKONIC CO., LTD.
7-24-14, Oizumigakuen-cho, Nerima-ku, Tokyo, Japan
Printed in Japan
Donated to www.orphancameras.comView original page 5From the M. Butkus archive. Original publisher and archive notices are retained.